Abstract
In quasi 3D mammography, it is necessary to have compressible phantoms that simulate breasts for specific quality controls and for research purposes. The materials used must have adequate density - X ray attenuation - radiological image - hardness. The objectives of this work were: to characterize the X ray beam in Molybdenum-Molybdenum quality and to obtain the attenuation coefficients of samples that simulate adipose, glandular, and malignant tissue. Half value layer measurements were performed at 22, 25, 28, 30, 32 y 35 kV on samples composed of silicones, gels, silicone oils, glass microspheres, quartz powder, polylactic acid and thermoplastic polyurethane. The samples of gel + Thinner + microspheres and gel + Thinner + quarts, for 15 keV, gave values of attenuation closer to adipose tissue. While the sample of gel + silicone + Thinner + microsphere could simulate malignant tissue.
Resumen
En mamografía cuasi 3D es necesario contar con fantomas compresibles que simulen mamas para la realización de controles de calidad específicos y para fines de investigación. Los materiales utilizados deben tener densidad - atenuación en rayos X - imagen radiológica - dureza adecuadas. Los objetivos de este trabajo fueron: caracterizar el haz de rayos X calidad Molibdeno-Molibdeno y obtener los coeficientes de atenuación de muestras que simulen tejido adiposo, glandular y maligno.
Se realizaron mediciones de capas hemirreductoras en 22, 25, 28, 30, 32 y 35 kV de muestras compuestas por siliconas, geles, aceites de silicona, microesferas de vidrio, polvo de cuarzo, ácido poliláctico y poliuretano termoplástico. Las muestras de gel + Thinner + microesferas y de gel + Thinner + cuarzo, para 15 keV, dieron valores de atenuación cercanos al del tejido adiposo. En tanto que la muestra de gel + silicona + Thinner + microesfera podrían simular tejido maligno.
About The Speaker
MsC. Rosana Pirchio
Discussion